The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. The head phantom designing involves slice-wise modeling of an adult head using PMMA. The phantom has provisions to hold detectors such as ionization chambers of different sizes, Gafchromic films, gel dosimeter, and optically stimulated luminescence dosimeter. For the point dose verification purpose, 15 volumetric modulated arc therapy patient plans were selected, and doses were measured using a CC13 ionization chamber. The percentage gamma passing rate was calculated for acceptance criteria 3%/3 mm and 2%/2 mm using OmniPro I'mRT film QA software, and Gafchromic EBT3 films were used for 2D planar dose verification. Treatment planning system calculated, and the measured point doses showed a percentage deviation ranged from 0.26 to 1.92. The planar dose fluence measurements, for set acceptance criteria of 3%/3 mm and 2%/2 mm, percentages of points having gamma value <1 were in the range of 99.17 ± 0.25 to 99.88 ± 0.15 and 93.16 ± 0.38 to 98.89 ± 0.23, respectively. Measured dose verification indices were within the acceptable limit. The dosimetric study reveals that head phantom can be used for routine pretreatment verification for the cranial RT, especially for stereotactic radiosurgery/RT as a part of patient-specific quality assurance. The presently fabricated and validated phantom is novel, versatile, and cost-effective, and many institutes can afford it. The dosimetric study reveals that head phantom can be used for routine pretreatment verification for the cranial RT, especially for stereotactic radiosurgery/RT as a part of patient-specific quality assurance. The presently fabricated and validated phantom is novel, versatile, and cost-effective, and many institutes can afford it. The purpose of this study is to evaluate the performance characteristic of volumetric image-guided dedicated-nozzle pencil beam-scanning proton therapy (PT) system. PT system was characterized for electromechanical, image quality, and registration accuracy. Proton beam of 70.2-226.2 MeV was characterized for short- and long-term reproducibility in integrated depth dose; spot profile characteristics at different air gap and gantry angle; positioning accuracy of single and pattern of spot; dose linearity, reproducibility and consistency. All measurements were carried out using various X-ray and proton-beam specific detectors following standard protocols. All electro-mechanical, imaging, and safety parameters performed well within the specified tolerance limit. The image registration errors along three translation and three rotational axes were ≤0.5 mm and ≤0.2° for both point-based and intensity-based auto-registration. https://www.selleckchem.com/products/liraglutide.html Distal range (R ) and distal dose fall-off (DDF) of 70.2-226.2 MeV proton beams were period of 8 months. The methodology and data presented here may help upcoming modern PT center during their crucial phase of commissioning. The PT-system performed well within the expected accuracy level and consistent over a period of 8 months. The methodology and data presented here may help upcoming modern PT center during their crucial phase of commissioning.This article explores the regulation of acupuncture in the UK and France. It focuses on the dilemmas such regulation has raised, and the effects of two contrasting approaches to the regulatory organisation of acupuncture within healthcare systems on practices and care. Although the question of how acupuncture, like other complementary, alternative or traditional therapies, should be regulated has often been reduced to a question of scientific knowledge, it is also dependent on the intricacies of national health system governance, state rationales and professional identities. France and the UK provide exemplary instances of contrasting systems, in which each of these factors has come to shape the regulation of the highly heterodox practice that is acupuncture. Overall, exploring the challenges of regulating acupuncture provides useful perspectives on how the make-up of legitimate therapies is constituted in particular healthcare contexts.The use of solid cooking fuels-wood, straw, crop residue, and cow-dung cakes-is associated with higher levels of environmental pollution and health burden. However, even in an era when incomes have grown and poverty has declined, the proportion of Indian households using clean cooking fuels such as kerosene or Liquefied Petroleum Gas (LPG) has increased only slightly. Even among the wealthiest quintile, only about 40 percent of the households rely solely on clean fuel. Since the chores of cooking and collection of fuel remain primarily the domain of women, we argue that intra-household gender inequalities play an important role in shaping the household decision to invest in clean fuel. Analyses using data from the India Human Development Survey (IHDS), a panel survey of over 41,000 households conducted in two waves in 2004-05 and 2011-12, respectively, show that women's access to salaried work and control over household expenditure decisions is associated with the use of clean fuel.This paper presents an individualized Ensemble Model Predictive Control (EnMPC) algorithm for blood glucose (BG) stabilization and hypoglycemia prevention in people with type 1 diabetes (T1D) who exercise regularly. The EnMPC formulation can be regarded as a simplified multi-stage MPC allowing for the consideration of N en scenarios gathered from the patient's recent behavior. The patient's physical activity behavior is characterized by an exercise-specific input signal derived from the deconvolution of the patient's continuous glucose monitor (CGM), accounting for known inputs such as meal, and insulin pump records. The EnMPC controller was tested in a cohort of in silico patients with representative inter-subject and intra-subject variability from the FDA-accepted UVA/Padova simulation platform. Results show a significant improvement on hypoglycemia prevention after 30 min of mild to moderate exercise in comparison to a similarly tuned baseline controller (rMPC); with a reduction in hypoglycemia occurrences ( less then 70 mg/dL), from 3.