https://www.selleckchem.com/products/p7c3.html To evaluate the intra and interobserver reproducibility of a new goniometric method for evaluating the isolated passive range of motion of the glenohumeral joint in an outpatient setting. This is a prospective observational study on healthy subjects. The Glenohumeral ROM Assessment with Scapular Pinch (GRASP) method is a new method for assessing the isolated range of motion (ROM) of the glenohumeral joint (GH) by a single examiner with a clinical goniometer. It measures the isolated glenohumeral passive abduction (GH-AB), passive external rotation (GH-ER) and internal rotation (GH-IR) with the arm at 45º of abduction. These three GH ROM parameters were measured in both shoulders of 30 healthy volunteers (15 males/15 females, mean age41.6[SD = 10.3] years). The full shoulder passive abduction, passive external rotation and internal rotation 45º of abduction were measured by the same examiners with a goniometer for comparison. One examiner made two evaluations and a second examiner made a third one. The primary outcome was the intra- and interobserver reproducibility of the measurements assessed with intraclass correlation coefficients (ICC) and the Bland-Altman plot. The intra-observer ICC for isolated glenohumeral ROM were 0.84 ± 0.07 for GH-ABD, 0.63 ± 0.09 for GH-ER, and 0.61 ± 0.14 for GH-IR. The inter-observer ICC for isolated glenohumeral ROM were 0.86 ± 0.06 for GH-ABD, 0.68 ± 0.12 for GH-ER, and 0.62 ± 0.14 for GH-IR. These results were similar to those obtained for full shoulder ROM assessment with a goniometer. The GRASP method is reproducible for quick assessment of isolated glenohumeral ROM. III. III. Four-rod instrumentation and interbody fusion may reduce mechanical complications in degenerative scoliosis surgery compared to 2-rod instrumentation. The purpose was to compare clinical results, sagittal alignment and mechanical complications with both techniques. Full spine radiographs were analysed in 97 pati